The HGC Humidity Generator and Controller is a unique and flexible solution for delivering an accurate relative
humidity environment for many applications. Especially useful for conditioning samples for laboratory and QA testing, the HGC can control the humidity of a chamber of up to nominally 2,000 ml volume to a relative humidity of 85%.
Principle of Operation:
Air is pumped through a desiccant built into the instrument. Automated
valving, controlled by a humidity sensor located at the sample
site, is used to divert some gas flow through a water reservoir
before being pumped to the sample. This feedback
function ensures that the humidity selected is the real humidity
at the sample site.
Features:
Simple software operation through Microsoft Excel™
Easy tie in with other equipment
Unique “sample site” humidity feedback control
Easy portability with compact design
Heated transfer line to avoid condensation
Applications include:
Thermogravimetry (TGA) |
Pharmaceuticals |
Atomic Force Microscopy (AFM) |
Foods |
Spectroscopy |
Polymers |
Thermo-Mechanical Analysis (TMA, DMA, DTMA) |
Biologics |
Rheology |
Petrochemicals |
Aerospace |
Defence |
Healthcare |
Forensics |
Medical |
Electronics |
Nuclear |
Ceramics |
Nanomaterials |
Many, many others |
Specifications:
Humidity Range: 5% to 90% (25°C)
10% to 85% (85°C)
Temperature Range: 5°C to 85°C*
NB: Take care with Dew Points for low
temperature studies
*Requires appropriate thermal chamber
Max. sample chamber volume: 2000 ml (Approx)
Instrument Footprint: 178mm depth x 365mm width x 350mm height
Instrument Weight: 14 kg
Connections: Electrical 85 to 264V AC, 150VA
Interface 1 USB input
Conformance:
Low Voltage Directive 73/23/EEC EN60950 1992
EMC Directive 89/336/EEC EN50081-1 1997
EN50082-1 1998
Conformity Mark CE 2003
Packing List:
Humidity Generator and Controller
2 x Humidity Sensor Assembly (calibrated)
Heated Line
Humidity Sensor Lead

The HGC is manufactured by Triton Technology Ltd.
Designs and specifications may change without notice.
E & OE
|

HGC Humidity Generator and Controller

HGC Programmed Humidity Control at Two Sample Temperatures

HGC "Staircase" Control of Humidity in Step and Ramp Profile
|