Features of the DSCUSB:
Quick and easy connectivity via USB
MODBUS RTU, MANTRABUS or ASCII protocols
Digital I/O
High speed, to 500 readings/second
Real mV/V calibration
Peak and trough recordings
No additional power supply required
Introduction
The DSCUSB is a compact, high performance digital signal conditioner with USB connectivity providing high resolution data for
portable applications. Simply by plugging the device into a PC, data
can be extracted from most strain gauge bridge input sensors and
acquired by software which allows data manipulation. No additional power supply is needed.
Although ideal for a 1 to 1
interface, multiple DSCUSBs may be connected to a single USB port via a suitable hub. A choice of MODBUS RTU, ASCII and MANTRABUS protocols at industrial stability are currently available.
Advanced Characteristics:
An Intelligent
Strain Transducer
High
Performance
1 Year
Manufacturer's Warranty
Serial
Communications - excellent stability
Fieldbus
connectivity - universal systems compatibility
Direct
output in Engineering Units if required
Resolution
selectable to 1 part in 66,000
"Plug
in and go" sensor - no need for a separate amplifier
No need
for an expensive A-to-D Card
Outstanding
performance-to-cost ratio
Five-point
transducer temperature compensation via comms
Seven-point
transducer linearisation via comms
Sensor
operation limit alarms
Reverse polarity protected
Continuous
auto-zero operation
Ideal
for precision measurement at low cost
Non-expert
maintenance
Self
diagnostics alerts user to common Strain Gauge faults:
Excitation short-circuits
Input over- or under-range
Cell scaling over- or under-range
System scaling over- or under-range
Temperature over- or under-range
Communication errors
Small
package size
Works
with most common strain gauge devices
Saves
cost and space
Simple
to use
Competitive
prices
Ideal
for OEM, R&D, Production, QA, QC applications
Full
CE approval
RoHS compliant
Options:
Encased version with Type A USB lead
PCB Version
DSJ2 Easy-connect module for PCB version
Precision vs Speed of Reading:
Generally, the higher the precision required from
an analogue-to-digital system, the slower the speed of conversion to digital
format. The DSCUSB allows you to select the resolution and speed of reading
that you require according to the Table below:
Speed
Readings/sec
(Hz) |
Resolution |
Divisions
1 part in... |
ppm |
%
of
Full Scale |
|
1
|
66,000
|
15
|
0.0015
|
|
10
|
40,000
|
25
|
0.0025
|
|
100
|
10,000
|
100
|
0.0100
|
|
500
|
5,000
|
200
|
0.0200
|
Choice
of Communication Options
&
Product
Order Codes:
|
DSCSUASC
|
Industrial Stability USB, ASCII Protocol, PCB version |
|
DSCSUEASC
|
Industrial Stability USB, ASCII Protocol, with enclosure
|
|
DSCSUMAN
|
Industrial Stability USB, Mantrabus Protocol, PCB version
|
|
DSCSUEMAN
|
Industrial Stability USB, Mantrabus Protocol, with enclosure
|
|
DSCSUMOD
|
Industrial Stability USB, Modbus Protocol, PCB version
|
|
DSCSUEMOD
|
Industrial Stability USB, Modbus Protocol, with enclosure
|
|
DSJ2
|
Junction PCB with Field Connectivity for DSCUXXX
|
Specifications:
| Description |
Units
|
Industrial
Stability |
| Min |
Typ |
Max |
| Bridge
Excitation: |
VDC
|
4.5
|
5
|
5.25
|
| Bridge
Impedence: |
Ohms
|
320
|
350
|
5,000
|
| Bridge
Sensitivity: |
mV/V
|
-3
|
+2.5
|
+3
|
| Offset
Temperature Stability: |
ppm/°C
|
-
|
5
|
10
|
| Gain
Temperature Stability: |
ppm/°C
|
-
|
30
|
50
|
| Offset
Stability With Time: |
%
FR*
|
-
|
0.0035
|
0.016
|
| Gain
Stability With Time: |
ppm
of FR/1st Year
|
-
|
-
|
300
|
| Nonlinearity: |
ppm of
FR
|
-
|
5
|
25
|
| Internal
Resolution: |
Counts/Divisions
|
-
|
16
Million
|
-
|
|
Resolution
@
1 Hz:
|
Counts/Divisions
over 100s
|
-
|
66,000
|
-
|
Resolution
@ 10 Hz: |
Counts/Divisions
over 100s
|
-
|
40,000
|
-
|
Resolution
@ 100 Hz: |
Counts/Divisions
over 100s
|
-
|
10,000
|
-
|
Resolution
@ 500 Hz: |
Counts/Divisions
over 100s
|
-
|
5,000
|
-
|
| Signal Filter : |
Dynamic recursive type, user programmable |
|
Optional:
Temperature Measurement Resolution
|
°C
|
-
|
0.0625
|
-
|
|
Optional:
Temperature Measurement Accuracy (-10 to 85°C)
|
°C
|
-
|
0.5
|
-
|
Optional:
Temperature Measurement Accuracy (-55 to 125°C) |
°C |
|
2.0 |
|
Optional:
Temperature Update Speed |
s |
- |
5 |
|
| Power
Supply Voltage: |
VDC
|
4.25
|
5
|
5.5
|
| Power
Supply Current (350R Bridge): |
mA
|
-
|
70
|
80
|
| Sensor
Impedence: |
Ohms
|
320
|
350
|
5,000
|
|
Excitation
System:
|
-
|
-
|
4-wire
|
-
|
| Operating
Temperature Range: |
°C
|
-40
|
-
|
+85
|
| Storage
Temperature: |
°C
|
-40
|
-
|
+85
|
| Humidity: |
%
RH Non-condensing
|
0
|
-
|
95
|
| Data Transmission Rate: |
kbps
|
2.4
|
-
|
460.8
|
| Output Cable Length (speed dependent) |
m
|
-
|
-
|
5
|
Dimensions:
Cased Version
|
86 x 57 x 26.5mm (3.35 x 2.25 x
1.9ā) excluding connector
(Length 95mm (3.74ā)
including 9-way āDā type socket)
Fitted with 1360 mm
(54") USB cable with Type A plug
Case material: extruded aluminium 6063
Bezels: polycarbonate
|
| Dimensions: PCB Version DSCUXXX |
43 x 28 x 12mm (1.693 x 1.103 x 0.473") |
Dimensions:
DSJ2 Junction PCB |
82 x 60 x 20mm (3.228 x 2.362 x 0.787") |
Environmental
Approvals:
|
EMC
Directive 89/336/EEC
Basic Standard BS EN 61326:1998
EMC Emissions
BS EN 55011:2007
EMC Immunity
BS EN 61000-42002
BS EN 61000-4-4:2004
BS EN 61000-4-11:2004:1995
BS EN 61000-4-3:2
Safety/Low Voltage Directive
73/23/EEC amended by 93/68/EEC
IEC 1010-1 :1990
BS EN 61010-1 :2001 |
* FR = Full Range
Downloads:
|

DSCUSB - Cased Version

Free Instrument Explorer and DSC Toolkit configuration, test and data display and data logging software
(Note: the load cell is not supplied, but we can supply a wide range of strain-gauge based load cells, pressure transducers and accelerometers)

DSCUSB PCB Only (top), and fitted in the DSJ2 Field Connectivity Module (bottom)
Related products:

DSC Digital Strain Cards
DCell Digital Strain Puck

ICA Analogue In-Cell
Strain Gauge Conditioning Amplifier

SGA Analogue Strain
Gauge Conditioning Amplifier
 
T24 Range High Speed Radio Telemetry
RoHS Statement
Buy securely OnLine now using

Please contact us for a quotation for quantity purchases

Please contact us for a Quotation
|